MODEL: | ECG 1200G |
Input Circuit | Floating; protection against defibrillator effect. |
Lead | Standard 12 leads |
Display | 8” TFT Display, 800 x 600 pixels, LED backlight |
Patient Current Leakage | 10µA |
Frequency response | 0.05Hz – 150Hz (-3dB) |
Input Impedance | 50MΩ |
Time Constant | >3.2seconds |
CMRR | >60dB, ≥100dB (Adding Filter) |
EMG Interference filter | 35Hz (-3dB) |
Recording | Thermal Printing System |
Specification of recording paper | 210mm x 30m roll paper, high speed paper |
Paper speed |
Auto record: 25mm/s, 50mm/s, error ± 5% Rhythm record: 25mm/s, 50mm/s, error ± 5% Manual record: 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s, error ± 5% |
Sensitivity selections | 5,10,20,40mm/mV, error± 5%. Standard sensitivity is 10mm/mV±0.2mm/mV |
Auto Record | Record setup according to auto record format and mode, automatically changing leads, measuring and analysing |
Rhythm Record | Record setup according to rhythm record format and mode, automatically measuring and analysing |
Manual Record | Record setup according to manual record format, manually changing leads |
Measurement parameters | HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc P axis, QRS Axis, T Axis R(V5), S(V1), R(V5) + S(V1) |
Product Safety Type | Class I CF applied part, there is defibrillation and pacing protection circuit in it |
Enduring Polarization voltage | ±500mV |
Power Supply AC | AC:100-240V, 50/60Hz DC: 14,8V, 3700mAh rechargeable lithium battery |
Noise Level | <15µVp-p |
Size | 334mm(L) x 320mm (W) x 85mm (H) |