| MODEL: | ECG 100G |
| Running Environment Temperature | ±5°C - +35°C |
| Relative humidity | ≤80% |
| Atmospheric pressure | 86kPa – 106kPa |
| Storage and transportation environment temperature | -10°C - 55°C |
| Relative humidity | ≤95% |
| Atmospheric pressure | 50kPa – 106kPa |
| ECG Input Circuit | Floating; protection against defibrillator effect. |
| Lead | Standard 12 leads |
| Patient Current Leakage | 10µA |
| Sampling Rate | 1000Hz |
| Frequency Response | 0.05Hz~150Hz (-3dB) |
| A/D change | 12 bites |
| Time constant: | >3.2 seconds |
| CMRR | >60dB >100dB (Add Filter) |
| Enduring polarization voltage | ±500mV |
| Noise Level | <15µVp-p |
|
Measurement parameters EMG Interference filter |
35Hz (-3dB) |
| AC filter | 50Hz/60Hz (≥20dB) |
| Sensitivity choice | 5,10,20mm/mV, error ±5% standard sensitivity is 10mm/mV ±0.2mm/mV |
|
Wave Output Recording way Auto-record Manual record Specification of recording paper Paper speed |
Thermal printing system Auto leads – changing Manual leads – changing 50mm (W) x 20mm (L) – high speed thermal paper 25mm/s, 50mm/s, error ± 5% |
| Product safety Type | Class I, type CF applied part, which has defibrillation – proof applied part |
| Fuse Specification | 2 pcs Ø5 x 20mm AC time lag: T250mA/250V (power supply: 220V) |
| Dimensions | 315mm (L) x 215mm (W) x 77mm (H) |
| Net Weight | 2.25kg |
| Power Supply |
AC: 220V, 50Hz (110V, 60Hz) DC: 7.4V, 3700 mAh rechargeable lithium battery |