PRODUCT CATEGORIES

AED HD1 / AED HD1VH1 / AED HD1VH2 PARAMEDIC SERIES MANUAL AND AED OPERATION DEFIBRILLATOR / MONITOR - CU-HD1





  • Efficient and effective e~cube Biphasic technology (BTE Type)
  • Manual and AED operation
  • Defibrillation using paddles, pads or internal paddles
Reference Printer ECG Cable SpO2 Pacer NIBP
 207-HD1-SP Yes 3 leads - - -
 207-HD1-VH1 Yes 3 leads Yes Yes -
 207-HD1-VH2 Yes 12 leads Yes Yes -
 207-HD1-VH3 Yes 12 leads Yes Yes Yes

  • Noninvasive pacing mode


SPECIFICATIONS
MODEL CU-HD1
Dimensions: 318mm(W) x 208mm (D) x 355mm (H)
Display: LCD Dimensions: 7 inch Diagonal (152mm * 91mm): 7 inch Diagonal (152mm * 91mm)
TFT Colour,   Resolution:   800 * 480 pixels
Defibrillation:
 
 
AED Mode:
 
 
 
 
 
Manual Mode:
Waveform:  Biphasic Truncated Exponential (e~cube)
Charge Time:  AC Power:  Less than 5 seconds to 200 Joules
Battery:  Less than 7 seconds to 200 Joules
Output Energy:  200J
Shock Delivery:  Via multifunction defib electrode pads
AED Develp Guideline:  Shock advisory sensitivity and Specificity meet AAMI DF-80 guidelines
Voice & Text Prompts:  Guide the user through the protocol Via Muftifunction defib electrode pads.
Out Put Energy: 1~10J,15J, 20J, 30J, 50J, 70J, 100J, 120J, 150J, 170J, 200J
Shock Delivery:  External paddle (with Pediatric), Internal paddle
Synchronous Cardioversion
Printer: Continuous ECG Strip:  Real-Time (8 seconds delay)
Auto Printing:  Recorder can be configured to print marked event, charge, shock and alarms,    Printing Speed:  25mm/s
                Paper:  50mm Width / 40 Diameters
Automatic Self-Test: Power on Self-Test
Run Time Self-Test
Manual Self-Test
Periodic Self-Test (Daily / Weekly / Monthly)
Power Source: External Battery Pack:  Lithium Polymer
Type:  14.8V 3.1Ah (Rechargeable)
Capacity:  When new, minimum of 100 shock deliveries (200J)
AC Power Module
Output:  18V, 5A
ECG Monitoring
Input:
 
 
 
 
 
Heart Rate Display:
ECG Size:
Heart Rate / Arrhythmia Alarm:
 
3-Lead Cable, I, II, III
5-Lead Cable, I, II, III, aVR, aVL, aVF or V
10-Lead Cable, I, II, III, aVR, aVL, aVF or V1, V2, V3, V4, V5, V6
(Display View:  All 12-Lead ECG waves display simultaneously)
Lead Fault:  “Lead Fault” message and dashed line display if an electrode or lead wire becomes disconnected
30 to 300bpm (±3bpm)
5, 10, 20mm/mV and Auto-gain
Heart Rate/Arrhythmia Alarm: Hr, Asystole, VF, VT
Noninvasive Pacing:
 
 
 

 
Waveform:  Monophasic Truncated Exponential
Mode:  Demand and Fixed Mode
Amplitude Accuracy:  0 ~ 200mA (± 5mA)
20ms (± 1.5%)
Pulse Width:  20ms (± 1.5%)
Pulse Rate: 30 ~ 180ppm (± 1.5%)
Refractory Period:  340msec (30 to 80ppm), 240msec (90 to 180ppm)
SPO2 Pulse Oximetry: Saturation: 70 ~100% (±3digits)
Pulse Rate: 20 ~250bpm (±3bpm)
Perfusion:  0.2%
Module Manufacturer:  Nellcor
SPO2 Alarm:  Less than Minimum setting rate
Over than Maximum setting rate
Data Storage: External Memory Card:  SD Card (ECG data, Event, Voice)
Standard Accessories:
  • External Paddles                        
  • Pads – 1 set adult
  • Pads Adapter
  • 3 or 12 Lead ECG Cable
  • Smart Card & Reader
  • SP02 Finger Probe with extension cable – Adult
  • 1 x pack of disposable electrodes
  • 1 x roll of printing paper
  • Battery Pack
  • Carrying Case
  • User Manual
Optional Accessories
  • 5 Lead ECG Cable
  • 10 Lead ECG Cable
  • Bed  Hanger
  • Pacer and SPO2 Module
  • Car Cigar Lighter Jack
  • SPO2 Finger Probe (Standard adult probe with VH1, VH2 and VH3 Model)
  • SPO2 Extension Cable (Standard Adult with VH1, VH2 and VH3 Model)