| Reference | Printer | ECG Cable | SpO2 | Pacer | NIBP |
| 207-HD1-SP | Yes | 3 leads | - | - | - |
| 207-HD1-VH1 | Yes | 3 leads | Yes | Yes | - |
| 207-HD1-VH2 | Yes | 12 leads | Yes | Yes | - |
| 207-HD1-VH3 | Yes | 12 leads | Yes | Yes | Yes |
| SPECIFICATIONS | ||
| MODEL | CU-HD1 | |
| Dimensions: | 318mm(W) x 208mm (D) x 355mm (H) | |
| Display: |
LCD Dimensions: 7 inch Diagonal (152mm * 91mm): 7 inch Diagonal (152mm * 91mm) TFT Colour, Resolution: 800 * 480 pixels |
|
|
Defibrillation: AED Mode: Manual Mode: |
Waveform: Biphasic Truncated Exponential (e~cube) Charge Time: AC Power: Less than 5 seconds to 200 Joules Battery: Less than 7 seconds to 200 Joules Output Energy: 200J Shock Delivery: Via multifunction defib electrode pads AED Develp Guideline: Shock advisory sensitivity and Specificity meet AAMI DF-80 guidelines Voice & Text Prompts: Guide the user through the protocol Via Muftifunction defib electrode pads. Out Put Energy: 1~10J,15J, 20J, 30J, 50J, 70J, 100J, 120J, 150J, 170J, 200J Shock Delivery: External paddle (with Pediatric), Internal paddle Synchronous Cardioversion |
|
| Printer: |
Continuous ECG Strip: Real-Time (8 seconds delay) Auto Printing: Recorder can be configured to print marked event, charge, shock and alarms, Printing Speed: 25mm/s Paper: 50mm Width / 40 Diameters |
|
| Automatic Self-Test: |
Power on Self-Test Run Time Self-Test Manual Self-Test Periodic Self-Test (Daily / Weekly / Monthly) |
|
| Power Source: |
External Battery Pack: Lithium Polymer Type: 14.8V 3.1Ah (Rechargeable) Capacity: When new, minimum of 100 shock deliveries (200J) AC Power Module Output: 18V, 5A |
|
|
ECG Monitoring Input: Heart Rate Display: ECG Size: Heart Rate / Arrhythmia Alarm: |
3-Lead Cable, I, II, III 5-Lead Cable, I, II, III, aVR, aVL, aVF or V 10-Lead Cable, I, II, III, aVR, aVL, aVF or V1, V2, V3, V4, V5, V6 (Display View: All 12-Lead ECG waves display simultaneously) Lead Fault: “Lead Fault” message and dashed line display if an electrode or lead wire becomes disconnected 30 to 300bpm (±3bpm) 5, 10, 20mm/mV and Auto-gain Heart Rate/Arrhythmia Alarm: Hr, Asystole, VF, VT |
|
|
Noninvasive Pacing: |
Waveform: Monophasic Truncated Exponential Mode: Demand and Fixed Mode Amplitude Accuracy: 0 ~ 200mA (± 5mA) 20ms (± 1.5%) Pulse Width: 20ms (± 1.5%) Pulse Rate: 30 ~ 180ppm (± 1.5%) Refractory Period: 340msec (30 to 80ppm), 240msec (90 to 180ppm) |
|
| SPO2 Pulse Oximetry: |
Saturation: 70 ~100% (±3digits) Pulse Rate: 20 ~250bpm (±3bpm) Perfusion: 0.2% Module Manufacturer: Nellcor SPO2 Alarm: Less than Minimum setting rate Over than Maximum setting rate |
|
| Data Storage: | External Memory Card: SD Card (ECG data, Event, Voice) | |
| Standard Accessories: |
|
|
| Optional Accessories |
|
|