| MODEL: | BM5 |
| Display | 10.4” colour TFT (800 x 600) |
| Dimensions | 23.8cm x 25cm x 16.3cm, Approx. 3.5kg |
| Display Parameters |
Up to 3 waves (ECG, SpO2, Respiration) Categorized alarms (3 priority levels) Visual Alarm lamp handle Heart rate tone Battery status External power LED |
| Interfaces |
DC input connector: 11 to 16 VDC, 3A max Defib Sync Output Signal Level: 0 to 5V pulse Pulse Width: 100 ± 10ms LAN digital output for transferring data VGA output |
| Battery |
Internal battery: sealed lead-acid Battery status indicator Operating Time: 2.5 hours typically (fully charged battery) |
| Thermal Printer |
Speeds:25, 50mm/sec Paper Width:58mm |
| Tabular Trends |
Memory Storage: 24 hours Date Interval:1, 5, 15, 30 min and 1 hour Tabular Format: One table for all variables |
| Graphical Trends | Display Duration: 30, 60, 90 min. 3, 6 , 12 hours |
| Standard Accessories |
3-lead patient cable 1 ea electrodes 10ea NIBP tubing, 3 m long 1ea Adult cuff 25-35cm, reusable 1ea SpO2 sensor extension cable 2 m) 1ea SpO2 sensor, reusable 1 ea DC adaptor, 18VDC, 2.5A 1ea |
| Optional Accessories |
Temperature probe Surface/Skin Reusable (Option) Thermal paper (Option) 5-lead patient cable (Option) |
| ECG | |
|
3 Leads, 5 Leads (Optional) 30 to 300bpm ± 3 bpm 0.5Hz to 40Hz 25mm/sec 0.5, 1, 2, 4 mV/cm Indicator on waveform display, user selectable ≥ 5MΩ ≥ 90dB at 50 or 60 Hz ± 5mVAC, ± 300mVDC ≤5s ≤8s VTAC/VFIB VTAC ASYSTOLIC |
| SpO2 | |
|
0% to 100% 30 to 300bpm 70% to 100% ± 2 digits 0% to 69% unspecified ± 3 bpm |
| NIBP | |
|
Oscillometric Manual: Single measurement Auto: Automatic intervals of 1,2,3,4,5,10,15,20,30min 1,2,4,8 hours 30 to 300mmHg Systolic: 60 to 250mmHg Mean Arterial Pressure: 45 to 235mmHg Diastolic: 40 to 220mmHg 40 to 200bpm |
| Respiration | |
|
5 to 120 breaths/min ± 3 breaths/min 25mm/sec |
| Temperature Unit | |
|
15ºC to 45ºC 25ºC to 45ºC ± 0.1ºC 15ºC to 24ºC ± 0.2ºC YSI Series 400 Temperature Probes |
| IBP | |
· Pulse rate Counting Range
|
2 -50 to 300mm Hg ± 1mmHg ± 1 digit (<100mmHg) ± 1% ± 1 digit (<100mmHg) 0 to 300 beats/min ± 200mmHg ±1mmHg ±1mmHg over 24 hours 5µV,V/mmHg |
| EtCO2 (Optional) | |
|
Infrared Absorption 0-99mmHg 0~ 150bpm ± 2mmHg, 0~40mmHg ± 5%, 41~76mmHg ± 10%, 77~99mmHg 1mmHg YES |