PMS6 PATIENT MONITOR S6





  • 10.4” Colour TFT display, CRT or larger display can be connected simultaneously.
  • AC Power and internal rechargeable battery
  • 13 kinds of arrhythmia analysis, HRV and S-T segment analysis
  • 60 seconds ECG waveform review
  • 7-Channel ECG waveforms displayed simultaneously
  • Drug dose calculation
  • Maximum 72-hour graphic and tabular trends of all Parameters
  • Adjustable 3 level audio and visual alarm
  • Anti-high-frequency electrosurgical equipment
  • Optional with Built-in 3 channels thermal recorder.  External Printer can be connected
  • Networked with central monitoring system by wired or wireless LAN
  • Applicable for Adult, Pediatric and Neonatal patient


MODEL: S6
Display 10.4” Diagonal Screen
640 x 480 Pixel Colour TFT
Displayed Parameters:
Time
Alarms
ECG
Heart Rate
NIBP
Pulse Oximeter
Respiration Rate
Trends
Temperatures
Trace Freeze
 
Battery-backed quartz crystal clock
High & low limits selectable on patient parameters
ECG Waveform Scale
Derived from ECG or SPO2
Pressure (systolic, mean and diastolic)
Pulse Rate, Pulse waveform and percent saturation
Respiration rate derived from ECG
HR, RR, NIBP, ST, TEMP, CO2 and SPO2
Two Channels
Traces A & B
ECG
Display Gain Scales
 
Sweep Speed
Input Resistance
Frequency Response
CMRR
Electrode Offset Potential
Baseline Recovery
Heart Rate Range
Resolution
Accuracy
Alarm
Lead off condition
Standard Lead, I, II, III, aVL, aVR, aVF, Vx
2.5mm/mV, 5mm/mV, 7mm/mV, 10mm/mV, 15mm/mV, 20mm/mV, 25mm/mV
6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
>5M Ohm (at 10Hz, not including patient cable)
0.05Hz – 100Hz (3dB)
>80dB
Maximum ±0.3V
<5s after 5KV defibrillation
15bpm – 300bpm
1bpm
<100bpm ±1% ≥100bpm ±2%
Heart rate high and low limits alarm delay <12s
Detected and displayed
ST Segment range
Accuracy
-1.2mV to +1.2mV
±0.02mV
SPO2
 
 
LED Specifications
 
 
Alarm Display
Display Update
Resolution
Range:  0% - 100%
Accuracy:  80% - 100%  ±2 digits, 70%-80% ±3digits, 0-69% unspecified
                     Wavelength         Radiant Power
RED             660±2nm               1.8mW   
IR                 905±10nm             2.0mW
SPO2 high and low limits alarm delay <7s
<5s
1%
Pulse Rate:
Measurement Range
Resolution
Accuracy
Alarm delay
 
20 to 250bpm
1bpm
±1bpm
Pulse rate high and low limits alarm delay <7s
NIBP
Technique
 
Patient Types
Cuff Inflation Time
Cuff inflation pressure
 
Measurement Modes
 
 
 
First Inflation Measurement
 
Oscillometric method (with inflatable cuff)
Determines systolic,diastolic and mean arterial pressures
Adult, Pediatric and Neonatal
3-15s depending on cuff size
Adult/Pediatric/Neonatal, subsequent inflation pressure determined by last measured systolic pressure
Manual:  Immediate upon operator command
AUTO:  Determinations automatically made with selectable intervals
STAT:  Determinations continues in 5 minutes
Adult/Pediatric/Neonatal
Measurement Interval Time 1 – 240min
Step:  1 min (1min – 10min), 5min (10 min – 30min)
          10 min (30min – 90min), 30min (90 – 240min)
Measurement Range Adult
Systolic: 30mmHg – 225mmHg
Diastolic:15mmHg – 220mm Hg
Mean Arterial:  20mmHg~235mmHg
Measurement Range Infant
Systolic:  30mmHg – 135mmHg
Diastolic: 15mmHg – 110mmHg
Mean Arterial:  20-125mmHg
Pressure Resolution:  1mmHg
Accuracy Cuff Pressure Range:  0 to 275mmHg
Pressure Span Accuracy:  ±3mmHg
Mean difference:  ±5mmHg
Standard deviation:  8mmHg
Determination Time Typically 25s.  Varies with patient’s pulse rate, pulse pressure and amount of artifact present
Overpressure Valve Automatically releases cuff pressure if inflation pressure exceeds 280mmHg/150mmHg
Overtime Protection Stop determinates if the measurement time exceeds 120s/90s
Alarm Delay Pressure high and low limits alarm delay <7s
RR
Technique
Range
Resolution
Accuracy
Alarm delay
 
Trans-thoracic impedance (RA-LL)
0-100 rpm
1 rpm
± 2 rpm (0-60rpm)
Respiration rate high and low limits alarm delay <7s
Temperature:
Technique
Channel
Scales
Probes

Range
Revolution
Accuracy
Alarm delay
 
Resistance
2 (T1 and T2)
F. or C.
Resistive, recta and skin (reusable and disposable)
YSI 400 Series types
0-50°C
0.1°C
±0.1°C
Body temperature high and low limits alarm delay <7s
CO2 (Optional)  
IBP (Optional)  
Power Requirements 100-240VAC, 50/60Hz  1.6A max
Internal Battery 12V 7AH, typed-sealed lead-acid
Operating Time 2 hours typical at 25°C, no printing, one NIBP measurement
(fully charged battery) Per 15 min