CMS9200 VITAL SIGNS MONITOR CMS9200





CMS9200 features in high cost-performance, high safety and low power consumption. To ensure monitoring quality and reduce operation risk, it adopts full-isolation(floating), ECG defibrillation-proof protection, ECG anti-high frequency surgical unit and NIBP dual-overpressure protection.

  • Display mode: 15'' color TFT LCD with high-resolution.
  • Be applicable for adult, pediatric and neonatal for all-round monitoring.
  • Operation interface with Chinese and English.
  • Waveform and data can be set optionally.
  • Storage of 72-hour trend data, and review of 40-second holographic waveform.
  • Full-lead ECG display.
  • 7-lead ECG displays in a screen, and ECG waveforms series display.
  • Anti-high frequency surgical unit, defibrillation-proof(requirement for special leads).
  • Adopt digital SpO2 technology, which has strong anti-interference and anti-weak filling capability.
  • Function of NIBP review, storage for up to 400 NIBP data.
  • Network: connecting with central station, other Bed observation and software updating.
  • Network connection mode: wireless and wired.
  • Built-in rechargeable battery for uninterrupted monitoring.
  • Standard parameters of ECG, RESP, NIBP, SpO2, TEMP and PR.
  • Built-in printer(optional).
  • Standard parameters of ECG, RESP, NIBP, SpO2, and TEMP.
  • IBP, CO2 and thermal printer are optional.



ECG
Lead Mode        3-lead and 5-lead are optional
Lead Selection  I, II, III, avR, avL, avF, V
Wave      5-lead: 2 channels
               3-lead: 1channel
Gain      ×0.25, ×0.5, ×1, ×2
HR Measuring and Alarm Range
Range     15 ~ 300 bpm
Accuracy     ±1% or ±1bpm, which is greater
Alarm Accuracy ±2bpm
Resolution 1 bpm
ST Segment Monitoring
Measuring and Alarm Range -0.6 mV~ + 0.8 mV
ARR
ARR Detecting Type    ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm Available
Review Available
Scan Speed for ECG Waveform   12.5mm/s, 25mm/s, 50mm/s
Respiration (RESP)
Method R-F(RA-LL) Impedance
Measuring and Alarm Range for Resp. Rate:
Adult 0~120rpm
Pediatric and Neonatal 0~150rpm
Resolution   1 rpm
Measuring Accuracy   2 rpm
Apnea Alarm    10 ~ 40 S
SpO2
Measuring and Alarm Range  0 ~ 100 %
Resolution   1 %
Accuracy:
                 70% ~ 100% ±2%
                 0% ~ 69% unspecified
PR
Measuring and Alarm Range  0~250bpm
Resolution   1bpm
Measuring Accuracy   ±2bpm or ±2%, which is greater
Alarm Accuracy   ±2bpm
NIBP
Method    Oscillometry
Mode    Manual, Auto, continuous
Measuring Interval in AUTO Mode 1 /2 /3 /4 /5/ 10/ 15/ 30 /60 /90 Min
Measuring Period in Continuous Mode 5 Min
Measuring and Alarm Range     10~270mmHg
Resolution   1 mmHg
Accuracy    ±10% or ±8mmHg, which is greater
Over-pressure Protection:           
         Adult Mode                     315±10 mmHg
         Pediatric Mode   265±10 mmHg
         Neonatal Mode   155±10 mmHg
TEMP
Channel    dual-channel
Measuring and Alarm Range  0 ~ 50C
Resolution   0.1C
Accuracy    ±0.1C
Optional
EtCO2
Method    Sidestream or Mainstream Measuring Range for CO2   0~150mmHg
Resolution for CO2:                                      
                             0.1 mm Hg  0 to 69 mm Hg  
                             0.25 mm Hg  70 to 150 mm Hg
Accuracy for CO2:
0 – 40 mm Hg  ±2 mm Hg  
41 – 70 mm Hg  ±5%  
71 – 100 mm Hg  ±8%  
101 – 150 mm Hg  ±10%  
Respiration Rate>80BPM ±12%
AwRR Range  2~150 rpm
AwRR Accuracy  ±1BPM
Apnea Alarm  Available
IBP
Channel   dual-channel
Label   ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and Alarm Range -50~350 mm Hg
Resolution  1 mm Hg
Accuracy   ±2% or 1mm Hg, which is greater
Recorder
 Print Mode thermal array
 Recording Width 48mm
 Paper Speed 25mm/s, 50mm/s
 Quantity of Printing Waveform dual-channel
Physical Characteristic:
Dimension 365 x 150 x 340 (mm)
Net Weight 4.0Kg
Accessories:
■Adult SpO2 probe(5-pin)
■Adult NIBP cuff
■Extending tube for NIBP
■ECG lead
■ECG electrode
■Temperature probe
■IBP sensor(optional)
■IBP integration module(optional)
■IBP integration module(optional)
Wall Bracket (optional)
■IBP cable(optional)
■CO2 module(optional)
■CO2 collection pipe(optional)
■Power cord
■User Manual